|
| |
Ludlum - Survey Meter
Operation
Ludlum Model 3
Ludlum Model 2241-2 Ludlum
Model 14C
|
Ludlum - Model 3
|

|
|
with 44-9 GM Pancake Probe
|
 |
|
The Ludlum Model 3 Survey Meter with Model 44-9 GM
Pancake Probe detects alpha, beta and measures gamma radiation levels form
0-200 mR/hr. It is typically used to measure lower levels of
radiation, and detects ion contamination levels. Its
size and shape provide easy handling for surveying personnel and
equipment. The pancake detector is intended to be used only
when surveying for contamination. The detector is housed in an
aluminum casing; the thin mica window of the detector is protected by a
stainless steel screen. The GM tube can be removed for replacement
if necessary. |
|
Suggested Operating Procedures - Ludlum Model 3 |
|
1. Attach the probe to the survey meter.
2. Insert fresh "D" cell batteries in the
instrument, observing polarity.
3. Check the battery strength by turning the selector
switch to the BAT position. The needle should jump to the
BAT TEST area.
4. Turn the selector switch to the x10 range, and
place the probe against a check source. The instrument should give
a reading.
5. Move the selector switch to view an
exponential reading.
6. Adjust the response control by flipping the toggle
switch to F (fast) or S (slow). The response should be adjusted to
get the most desirable compromise between response speed and meter fluctuation. |
|
|
|

|
|
|
Ludlum Model 2241-2 Survey Meter |
|
The Model 2241-2 is a portable microprocessor-based
digital Scaler/Ratemeter designed for use with a variety of detector
types for measurement of ionizing radiation. The data is presented
on a 4-digit (6 digits while in the scaler mode). Liquid Crystal
Display (LCD) with a moving decimal point. A three-position switch
labeled OFF/RATEMETER/SCALER selects the desired operating mode
for the instrument. A locking toggle switch labeled DET 1
and DET 2 allows for a quick change of operating parameters between
detectors. |
|
The Model 2241-2 features adjustable alarm set points for the
RATEMETER and SCALER operating mode. The RATEMETER mode has two
alarm indications. The first level alarm is indicated by the word
ALERT on the LCD. The second level alarm is indicated by by the
word ALARM and a continuous audible tone. The SCALER alarm
condition will also display the word ALARM and produce the same audible
tone. Both of the audible alarms can be silenced/acknowledged by
depressing the RESET switch. Alarm parameters are set during
instrument calibration.
The Unit is operated with two-"D" cell flashlight
batteries. The unit body is made of cast and drawn aluminum with
polyurethane paint which aids in decontamination of surfaces. |

|
|
Detector Types: |
|
The Ludlum Model 2241-2 can operate with a variety of
detectors, but will most commonly be used with the Model 44-9 (pancake)
and Model 44-38 (hotdog) external GM detectors. |
|

Model 44-9 (Pancake) Probe |
|
Instrument Operating Range: |
|
The instrument provides auto-ranging form 0 uR/hr to 9999 R.hr, or 0 cpm
to 999 kcpm, depending on which detector is attached. (u=micro = one
one-millionth; k-kilo = one thousand)
|
|
Types of Radiation Detected |
|
When operated with the Model 44-9 (pancake/frisker) detector, the
instrument is capable of detecting alpha, beta, gamma, and x-ray
radiation. When operated with the Model 44-38 (hotdog) external GM
detector, the instrument is capable of detecting beta, gamma, and x-ray
radiation.
The size and shape of the 44-9 provides easy handling
for surveying personnel and equipment. The pancake detector
is intended to be used only when surveying for contamination. The
detector is housed in an aluminum casing; the thin mica window of the
detector is protected by a stainless steel screen. The GM tube can
be removed for replacement if necessary.
|
|

Model 44-38 (Hotdog) Probe |
|
The Ludlum Model 44-38 GM (Hotdog) Detector will detect beta and gamma
radiation. The hotdog detector is intended to be used when
surveying for direct radiation versus contamination. The GM tube
is protected by a slotted metal housing assembly. The detector
housing incorporates a rotary shield which may be ued to differentiate
between beta and gamma radiation. With the detector shield closed,
beta radiation is blocked out and only gamma radiation is
detected. With the rotary shield open, both beta and gamma
radiation can be detected. The beta dose contribution from a
measurement can be determined by subtracting the reading taken with the
rotary shield closed from the reading takein with the rotary shield open
( open window closed window). |
|
Operating Controls: |
|
Off/Ratemeter/Scaler Switch: A three-position
rotary switch which applies power to the instrument and selects the
RATEMETER or SCALER counting modes. DETector Select
Switch: A protected toggle switch that toggles between DET 1
and DET 2 parameters. The pull-and-change toggle switch is
designed to protect against inadvertent changes. The switch is to
be set to DET 1 when the pancake probe is attached; units
displayed will be kcpm. Set the switch to DET 2 when the hotdog
probe is attached; units displayed will auto range from uR/hr to mR/hr
to R/hr. AUD ON/OFF Switch: The audio may
be silenced or enabled via this toggle switch. The audible ALARM
is independent of the AUD ON/OFF switch. F/S
(Fast/Slow) Response Switch: A two position toggle switch with
selects Fast or Slow counting response time. The fast response
setting should be used when performing an initial survey or when trying
to locate sources of radiation. Switch to slow response when more
accurate measurements are needed ( e.g. for official measurements). Light
(LCD Backlight): A pushbutton switch, when depressed,
illuminates the LCD for a pre-programmed period of time. RESET
Pushbutton Switch: In the non-alarm condition, depressing the
RESET switch resets the Ratemeter display to the minimum display
readout. In an alarm condition, depressing the RESET will silence
the audible alarm; depressing the RESET Switch a second time will reset
the Ratemeter alarm and/or alert condition. The Scaler alarm
can only be reset by depressing the Scaler to start the counting
cycle. Scaling time is set during instrument calibration and can
be set from 1 to 999 seconds. Depressing the Count switch after a
Scaler ALARM will reset the Scaler display to "0" resetting
the alarm condition. |

|
|
Suggested operating Procedure for the Ludlum Model
2241-2
|
|
1. Check the Calibration expiration date, if the instrument calibration is
past due, do not use the instrument.
2. Visually inspect the instrument for damage. If there is damage
which impairs the use of the instrument do not use the instrument.
3. Ensure the Model 2241-2 power switch is in the OFF position.
Open the battery lid and install two "D" size batteries.
Note (+) (-) marks on the inside of the lid. Match the battery
polarity to these marks. Close the battery box lid.
4. Connect either the hotdog or pancake detector to the
instrument and position the DETector select switch accordingly.
5. Turn the OFF/RATEMETER/SCALER switch to the RATEMETER
position. The display will go through a start-up sequence. The
display will show all 8s with decimal points. (see figure at
right) Check to make sure all segments of each number display are
functioning.
6. After the start-up sequence is complete, the unit will show the
typical operating screen with the display auto ranging to the current
radiation level in either R/hr or kC/m, depending on which position the
DETector select switch is in.
7. Expose the detector to an external check source if the background
count is too low to generate a display reading. Switch the AUD
ON/OFF to the ON postion and confirm the external speaker produces and
audible click. Normal operating model should be with the AUD ON/OFF
toggle switch in the ON position.
8. Switch to the Scaler position. Depress the Count switch
located in the ind of the carrying handle to start the count cycle.
The word "COUNTING" should be visible in the display during the
count cycle and should disappear at the end of the preset count
time. Count time is set during instrument calibration, ans is
typically set for six seconds.
9. Depress and release the LIGHT switch. The backlight
located behind the LCD should illuminate.
10. Select the desired F/S, AUD ON/OFF, and RATEMETER or SCALER
parameters and proceed to use instrument.
|
|
|
|
|
Ludlum Model 14C - Survey Meter
|
|
The Ludlum Model 14c is designed to be operated with an external GM
detector for X0.1 through X100 ranges and is equipped with an internal GM
detector for X1000 range.
The instrument provides an overall range of 0-2000 mR/hr
|
|
|
Types of Radiation detected.
|
|
Depending on which external GM detector is attached, the instrument is
capable of detecting alpha, beta, gamma and x-ray radiation. The
internal GM detector, operated on the X1000 scale only, will detect gamma
and x-ray radiation.
|
|
Operator-adjustable controls
|
|
The range mulitplier selector switch is a 6-position marked OFF,
X1000,X100,X10,X1X0.1. Moving the range selector switch to one of
the range mulitpolier postions provides the operator with an overall range
of 0-2000 mR.hr. Multiply the scale reading by the mulitiplier for
determining the actual scale readings.
Audio ON-OFF toggle switch: in the ON position, the switch
energizes the speaker, located on the left side of the instrument.
The frequency of the clicks is relative to the rate of the incoming
pulses. The higher the rate, the higher the audio frequency.
The audio should be turned OFF, when not required, to reduce battery drain
Fast-Slow toggle switch: selecting the fast, "F",
position of the toggle switch provides 90% of final meter reading in four
seconds. In slow, "S", position, 90% of final meter
reading takes 22 seconds. Set on "F" for fast response and
large meter deviation. "S" position should be used for
slow response and damped meter deviation. The fast response setting
should be used when performing an initial survey or when trying to locate
sources of radiation. Switch to slow response when more accurate
measurements are needed (e.g, for official measurements).
RES button: when depressed, provides a rapid means todrive the
meter to zero.
BAT check: when depressed, provides a visual means of checking
the battery charge status. The instrument must be turned on to
perform this function.
|
|
|
Operation of internal/external detector:
|
|
The internal GM detector is operational on the X1000 scale only. The
attached external GM detector is operational for scale multiples X0.1, X1,
X10, X100.
|
|
Specific characteristics/limitations:
|
|
The Ludlum Model 14C is a portable survey instrument designed for
operation with a GM detector that operates on two standard
"D" cell batteries. The unit body is made of
cast-aluminum, including the meter housing. The instrument can
operate with either a thin wall GM detector or a pancake GM
detector. The Ludlum Model 14c is also equipped with an internal
high range detector that is operational only when the meter is on the
X1000 range.
With two "D" cell batteries, the Ludlum Model 14c is set for
900-volt GN tube operations from 32 F to approx. 150 F, either very fresh
alkaline batteries or rechargeable Nickel-cadmium (NiCd) batteries may be
used.
|
|
Instrument Maintenance:
|
|
Never store the instrument over 30 days without removing batteries.
Although the instrument will not be affected, battery failure can occur.
Follow your state's requirements for instrument calibration,
Annual instrument calibration is recommended.
|
|
Suggested Operating Procedure for the Ludlum Model
14c:
|
|
1. Check the calibration expiration date, if the instrument calibration
is past due, do not use the instrument.
2. Visually inspect the instrument for damage. If there is damage
which impairs the use of the instrument do not use the instrument..
3. Ensure the Model 14c power switch is in the OFF postion. Opern the
battery lid and install two "D" size batteries. Note (+)(-) marks on the insde of the lid. Match the battery
polarity. If the response falls below the accepted range, install
new batteries.
4. Turn the instrument range switch to X1000. Depress the
button on the face of the instrument labeled BAT. The meter should
deflect to the battery check position of the meter scale. If the
meter does not respond, recheck to be certain that the batteries have
proper polarity. If the response falls below the accepted range,
install new batteries.
5. Utilizing the externally attached detector, move the range switch to
the lower scales until a meter reading is indicated. The toggle
switch labeled F-D should have fast response in "F" position,
slow response in "S" position. . Place the toggle switch
in the "F" position to start. Expose the external probe to
a radiation check source. The speaker should click with the audio
switch in the ON position. Visually inspect to ensure that the
readout of the instrument also responds.
6. Press the RES switch, The meter should zero.
7. Proceed to use the instrument.
|
|
|
|