Testing and Calibration

We are currently Testing and Calibrating Direct Reading Dosimeters, Electronic Dosimeters and select models of Survey Meters and Area Monitors

To return products to us for repair / calibration please fill out this form ( Repair & Calibration Request Form)  and return it with your shipment, this form also contains a Credit Application for those desiring credit terms with us.

Step 3 of tour Testing and Calibration

Arrow-Tech, Inc.  does the testing and calibration of our own dosimeters after they are manufactured.
We also can test and calibrate all makes and models of Direct Reading Dosimeters you may haveContact us for pricing and any special needs you may have.

Our Dosimeter calibration consists of the 4 following tests:

  •   Electrical Leakage

  •  Hermetic seal

  •  Mid-Scale Accuracy test

  •  Ruggedness test (Arrow-Tech dosimeters only)

  • Our testing procedures meet ANSI N322-1997 standards

  • NIST Traceability

  • Cs - 137 source is used for the testing.

  • Complete report on failed and passed dosimeters provided.

  • We can design a testing program to mach you licensing  requirements

Electrical Leakage 

Dosimeters are set to zero and are allowed to stand for 1 hour, recheck readings and placed in 50 degree Celsius oven for 5 days.   They are removed from oven and allowed to cool to room temps.   The Dosimeter is read and recorded. Any dosimeter that has changed by more than 1.25.5%/Day  (ANSI 2.5%/Day) of full-scale reading is rejected.


Hermetic  Seal    

Dosimeters are set to zero and placed on a rack in bell jars andVacuum_testing.jpg (41332 bytes) vacuum pumped for 48 hours (vacuum pressure 75 micro inches of mercury).  Then the bell jar is placed  in vacuum source and exposed to a mid scale  radiation dose.  The dosimeter is removed from bell jars and the reading is recorded. Any dosimeter reading less than 90% of  Reference Radiation reading is rejected.    A record is kept of  whether the dosimeters passed or failed test.
Gail starting the Mid Scale test using the 100 curie radiation source.   The Dosimeters are given a predetermined amount of radiation to read Mid scale.  For example the range being checked here is the 600 R.  They are being exposed for 20 minutes.  The reading on the pocket dosimeter should read 300R.  Those dosimeters that read within  + or - 10% are acceptable.   Those outside that range are failed. (Customer calibration)


Accuracy Test   

Dosimeter are set to zero and placed in  a carousel, then exposed to a mid-scale radiation dose.   The readings are recorded and any dosimeter that exceeds +/ - 8% (ANSI +/- 10%) of the mid-scale reading is rejected. Picture shows the 20 curie radiation source inside source room.

20 curie Cs-137 source -  For testing low range dosimeters. This was manufactured by J L Shepherd & Associates.

 One of Two- 100 curie Cs-137 source's - For testing high range dosimeters. This was designed and manufactured by the Oak Ridge National Labs. especially for Arrow-Tech, Inc.

Each source room has a 2 inch thick lead door to prevent any radiation from escaping during testing.

Monitoring station's to view what is being tested inside the source room. There is a monitor for each source room.
Alarm system in radiation source room


Ruggedness Test

Pat Jeannotte starting the ruggedness test by pouring the dosimeters down the stair step apparatus.  This shakes any dust or dirt particles that may be present after the manufacturing process.  If any particles are visible when looking through the dosimeter, the dosimeter is discarded.


  Back to Radiation Detection    Tour  Next step 4 of tour
 